Thesis
Pembuatan Desain Geometri Bur Implan Gigi Prototipe
PEMBUATAN DESAIN GEOMETRI
BUR IMPLAN GIGI PROTOTIPE
Erwin Albertus Launardo - 160521190001
ABSTRAK
Latarbelakang: ...
-
Code CallNo Lokasi Ketersediaan TH 963 963 617.63 Lau P Sekeloa (Periodonsia) Tersedia namun tidak untuk dipinjamkan - No Loan -
Perpustakaan Fakultas Kedokteran GigiJudul Seri -No. Panggil 963 617.63 Lau PPenerbit PPDGS Periodonsia FKG Unpad : FKG Sekeloa Unpad., 2022 Deskripsi Fisik xxxvii, 56 hlm, ilus; 21x30 cmBahasa IndonesiaISBN/ISSN 160521190001Klasifikasi 617.63Tipe Isi -Tipe Media -Tipe Pembawa -Edisi -Subyek Info Detil Spesifik -Pernyataan Tanggungjawab Erwin Albertus Launardo (Author) -
PEMBUATAN DESAIN GEOMETRI
BUR IMPLAN GIGI PROTOTIPE
Erwin Albertus Launardo - 160521190001
ABSTRAK
Latarbelakang: Perawatan implan gigi saat ini menjadi perawatan gigi alternatif
terbaik. Namun harga implan gigi tergolong mahal karena implan gigi masih di
impor dari luar negeri. Departemen Periodonsia Fakultas Kedokteran Gigi
Universitas Padjadjaran bekerjasama dengan ITB dan PT Pudak Scientific
berencana mendesain dan memproduksi fixture implan dalam negeri beserta dengan
bur implan gigi yang ideal untuk prototipe fixture implan yang telah berhasil
diproduksi. Tujuan: mengidentifikasi desain geometri bur implan komersial dan
menganalisa desain geometri bur implan dengan metode scoping review. Kemudian
membuat DED bur implan gigi yang baik dalam menjaga perubahan temperatur
saat osteotomi dan mengoptimalkan waktu pengeburan. Metode penelitian:
analisis geometri dua bur implan komersial dengan menggunakan mikro-CT,
pengkajian artikel mengenai desain bur implan gigi dengan metode scoping review,
pembuatan desain geometri bur secara tiga dimensi dan Detailed Engineering
Design (DED). Hasil: analisis dua bur implan komerisal memberikan ukuran
geometri yang berbeda. Ekstraksi data metode scoping review mendapatkan total
17 artikel yang membahas desain geometri bur implan dengan parameter seperti
temperatur pengeburan dan waktu pengeburan. Desain bur prototipe implan gigi
dibuat dengan menyesuaikan spesifikasi fixture implan merah putih. Simpulan:
Desain geometri bur prototipe implan gigi berdasarkan analisis geometri bur implan
komersial dan hasil kajian scoping review dibuat dengan diameter 3,4mm; point
angle 118,86o
dan 19,26o
, helix angle 0
o
, clearance angle 66,24o
, 3-flute, margin
bur yang besar pada badan bur serta desain rake face plane pada cutting lip.
Kata Kunci : implan gigi, desain bur tulang, nekrosis tulang, waktu pengeburan
v
GEOMETRY DESIGN OF
A PROTOTYPE DENTAL IMPLANT DRILL
Erwin Albertus Launardo - 160521190001
ABSTRACT
Background: Dental implant treatment is currently the best alternative to dental
restoration. However, the price of dental implants is quite expensive because dental
implants are still imported from abroad. The Department of Periodontology,
Faculty of Dentistry, Padjadjaran University, in collaboration with ITB and PT
Pudak Scientific, plans to design and produce domestic implant fixtures along with
dental implant burs that are ideal for the prototype implant fixtures that have been
successfully produced. Aim: to identify commercial implant drill geometry and to
analyze bone drill geometry design according to the result of the scoping review
methods. Then make dental implant drills DED that are good at keeping
temperature changes during osteotomy and optimizing drilling time. Methods: The
design of the dental implant prototype drill geometry is based on the geometric
analysis of commercial implant drills and the result of a scoping study, manufacture
of three-dimensional drill geometry designs and Detailed Engineering Design
(DED). Result: analysis of two commercial implant drills gives different geometric
sizes. Data extraction using the scoping review method yielded a total of 17 articles
discussing the geometrical design of bone drill with parameters such as drilling
temperature and drilling time. The design of the dental implant prototype drill was
made according to the specifications of the implant fixture prototype. Conclusion:
The prototype drill geometry design for dental implants based on the results of a
scoping review study was made with a diameter of 3.4mm; point angle 118,86o
and
19,26o
, helix angle 0o
, clearance angle 66.24o
, 3-flute, large portion of drill margins
on and rake face plane design on the cutting lip.
Keywords: dental implant, bone drill design, osteonecrosis, drill time -
Tidak tersedia versi lain
-
Silakan login dahulu untuk melihat atau memberi komentar.
//