Daya yang Dihasilkan Pegas Tork untuk Perawatan Ortodonti
Analysis of data obtained justifies the following conclusions :
1. Torqueing spring with one spur, 2cm legs made form 0,016 round wire exerted ...
-
Code CallNo Lokasi Ketersediaan Ort1-068A 68A 617.6 Ham D Jatinangor (Orthodonti) Tersedia namun tidak untuk dipinjamkan - No Loan -
Perpustakaan Fakultas Kedokteran GigiJudul Seri -No. Panggil 617.6Penerbit FKG Unpad : ., 1988 Deskripsi Fisik -Bahasa IndonesiaISBN/ISSN -Klasifikasi OrthodontiTipe Isi -Tipe Media -Tipe Pembawa -Edisi -Subyek Info Detil Spesifik -Pernyataan Tanggungjawab - -
Analysis of data obtained justifies the following conclusions :
1. Torqueing spring with one spur, 2cm legs made form 0,016 round wire exerted too much force over 300 grams.
2. Torqueing soring with one spur, made from 0,014 round wire with 2,5cms legs exerted 302 gram force maximally.
3. The longer legs made into the spring, the lower force exerted.
4. The thicker wire used, the higher force exerted.
5. The more spurs used, the lower force exerted.
6. Percentage of forece loss was considerably low after 2 weeks activated (6,4%). -